Q. How do I become involved with CRAG? A. E-mail CRAG secretary
Q. How do I configure my GE Innova 2000/2100 to produce the lowest possible dose settings? A. On the DL screen select Service > Configuration> Application Customization> Auto exposure preference> Set this to "Receptor Dose Limited Standard". This will give you approximately 30% reduction in dose over "IQ Standard".
Q. How do I found out more about the generic worker (Cardiac Catheter Lab Practitioner) course? A. For more information contact
Q. Which is the best equipment to choose when starting a new Lab? A. Download the latest "Cardiovascular imaging systems: a comparative report" from the NHS purchasing and supply agency website. If you are still unsure after your site visits and want an unbiased opinion - contact us.
| |||||||||